This compact and versatile temperature/humidity data logger is engineered for multi-use monitoring in transportation, storage, production, and HVAC environments. With a basic accuracy of T±0.3°C, RH±2.5%rh (20%rh to 80%rh) or optional T ±0.2°C, RH ±2.0%rh (20%rh to 80%rh) and IP56-rated protection.
The device is easily configurable using PoChain software, which is available for download on the manufacturer’s website. Users can set the logger to begin recording immediately or with a delayed start, depending on operational needs. Once data logging is complete, the unit connects directly to any computer via USB and automatically generates a compliant PDF report—no software installation required for basic report access.
For in-depth data analysis and customized export, users can connect the device to a computer running PoChain software. The software supports exporting data in multiple formats, including XLS, CSV, PDF, TXT, and CIN, allowing for smooth integration with record-keeping systems and quality assurance documentation.
The logger supports up to 6 data marker points, which can be manually triggered during the recording process. Pressing the key in mark mode (MK) allows users to flag specific events or time points for easier review and correlation later in the analysis phase.
Each unit includes a calibration certificate from an ISO/IEC 17025:2017 accredited lab, with full traceability to SI units through NIST. For long-term use, the device can be re-calibrated through PoChain software at up to 6 calibration points for each channel. This process should be performed by qualified laboratories—either by STN or any facility with proper calibration capabilities—to maintain accuracy and data integrity over time.
This high-performance temperature/humidity logger complies with food safety and pharmaceutical industry standards, making it an excellent choice for cold chain monitoring, warehouse storage validation, and temperature-sensitive production control.
Probe type |
Internal/External RH+T sensor |
Unit |
°C, °F & %rh |
Resolution |
0.1°C & 0.1%rh |
Measurement range |
-40°C to 70°C & 0.1%rh to 99.9%rh |
Accuracy |
Basic T: ±0.3°C, RH: ±2.5%rh (20%rh to 80%rh), ±3.5%rh (else) |
Report generation |
Automatic pdf report generation |
Export file formats |
pdf, txt, xls, csv, cin |
Data storage capacity |
25920 groups |
Memory |
Non-volatile memory, no data loss when power off |
LCD refresh interval |
10 seconds |
LCD display auto-off time |
3 minutes |
Sampling interval |
User configurable from 10 seconds to 18 hours |
High and low alarm values |
User configurable |
Alarm delay |
User configurable from 0 to 960 mins |
Alarm type |
Single time or cumulative, user configurable |
Start delay |
User configurable from 0 to 254 mins |
Start recording mode |
Press and hold the button for about 3 secs |
Stop recording mode |
Press button to stop; Automatic stop when the memory is full; When connected to a computer |
Operation/storage temperature |
-40°C to 70°C |
Communication interface |
USB 2.0 |
PC Software |
Software PoChain (Windows 7/8/Vista/10 32/64bits) website downloadable |
Power supply |
3.6V lithium battery, user replaceable |
Battery life |
About 3 years @15 minutes sampling interval, 25°C |
Protection Class |
IP56 (int sensor)/IP40 (ext sensor) |
Shell material |
ABS |
Dimensions/weight |
120mm*37mm*23mm/67g |
Certification and standards |
CE | ROHS | EN12830 | REACH |
Calibration |
Max. 6 points calibration by software |
Calibration report |
Calibration report included with each unit |
Test points |
20%rh, 50%rh and 80%rh @ 25°C and 25°C on the report |